Application of Mixed Effects Limits of Agreement in the Presence of Multiple Sources of Variability: Exemplar from the Comparison of Several Devices to Measure Respiratory Rate in COPD Patients.

Application of Mixed Effects Limits of Agreement in the Presence of Multiple Sources of Variability: Exemplar from the Comparison of Several Devices to Measure Respiratory Rate in COPD Patients.
Author Information (click to view)

Parker RA, Weir CJ, Rubio N, Rabinovich R, Pinnock H, Hanley J, McCloughan L, Drost EM, Mantoani LC, MacNee W, McKinstry B,


Parker RA, Weir CJ, Rubio N, Rabinovich R, Pinnock H, Hanley J, McCloughan L, Drost EM, Mantoani LC, MacNee W, McKinstry B, (click to view)

Parker RA, Weir CJ, Rubio N, Rabinovich R, Pinnock H, Hanley J, McCloughan L, Drost EM, Mantoani LC, MacNee W, McKinstry B,

Advertisement
Share on FacebookTweet about this on TwitterShare on LinkedIn

PloS one 2016 12 1411(12) e0168321 doi 10.1371/journal.pone.0168321
Abstract
INTRODUCTION
The Bland-Altman limits of agreement method is widely used to assess how well the measurements produced by two raters, devices or systems agree with each other. However, mixed effects versions of the method which take into account multiple sources of variability are less well described in the literature. We address the practical challenges of applying mixed effects limits of agreement to the comparison of several devices to measure respiratory rate in patients with chronic obstructive pulmonary disease (COPD).

METHODS
Respiratory rate was measured in 21 people with a range of severity of COPD. Participants were asked to perform eleven different activities representative of daily life during a laboratory-based standardised protocol of 57 minutes. A mixed effects limits of agreement method was used to assess the agreement of five commercially available monitors (Camera, Photoplethysmography (PPG), Impedance, Accelerometer, and Chest-band) with the current gold standard device for measuring respiratory rate.

RESULTS
Results produced using mixed effects limits of agreement were compared to results from a fixed effects method based on analysis of variance (ANOVA) and were found to be similar. The Accelerometer and Chest-band devices produced the narrowest limits of agreement (-8.63 to 4.27 and -9.99 to 6.80 respectively) with mean bias -2.18 and -1.60 breaths per minute. These devices also had the lowest within-participant and overall standard deviations (3.23 and 3.29 for Accelerometer and 4.17 and 4.28 for Chest-band respectively).

CONCLUSIONS
The mixed effects limits of agreement analysis enabled us to answer the question of which devices showed the strongest agreement with the gold standard device with respect to measuring respiratory rates. In particular, the estimated within-participant and overall standard deviations of the differences, which are easily obtainable from the mixed effects model results, gave a clear indication that the Accelerometer and Chest-band devices performed best.