Bedside Glucose Monitoring-Is it Safe? A New, Regulatory-Compliant Risk Assessment Evaluation Protocol in Critically Ill Patient Care Settings.

Bedside Glucose Monitoring-Is it Safe? A New, Regulatory-Compliant Risk Assessment Evaluation Protocol in Critically Ill Patient Care Settings.
Author Information (click to view)

DuBois JA, Slingerland RJ, Fokkert M, Roman A, Tran NK, Clarke W, Sartori DA, Palmieri TL, Malic A, Lyon ME, Lyon AW,


DuBois JA, Slingerland RJ, Fokkert M, Roman A, Tran NK, Clarke W, Sartori DA, Palmieri TL, Malic A, Lyon ME, Lyon AW, (click to view)

DuBois JA, Slingerland RJ, Fokkert M, Roman A, Tran NK, Clarke W, Sartori DA, Palmieri TL, Malic A, Lyon ME, Lyon AW,

Advertisement
Share on FacebookTweet about this on TwitterShare on LinkedIn

Critical care medicine 2017 02 06() doi 10.1097/CCM.0000000000002252
Abstract
OBJECTIVES
New data have emerged from ambulatory and acute care settings about adverse patient events, including death, attributable to erroneous blood glucose meter measurements and leading to questions over their use in critically ill patients. The U.S. Food and Drug Administration published new, more stringent guidelines for glucose meter manufacturers to evaluate the performance of blood glucose meters in critically ill patient settings. The primary objective of this international, multicenter, multidisciplinary clinical study was to develop and apply a rigorous clinical accuracy assessment algorithm, using four distinct statistical tools, to evaluate the clinical accuracy of a blood glucose monitoring system in critically ill patients.

DESIGN
Observational study.

SETTING
Five international medical and surgical ICUs.

PATIENTS
All patients admitted to critical care settings in the centers.

INTERVENTIONS
None.

MEASUREMENTS AND MAIN RESULTS
Glucose measurements were performed on 1,698 critically ill patients with 257 different clinical conditions and complex treatment regimens. The clinical accuracy assessment algorithm comprised four statistical tools to assess the performance of the study blood glucose monitoring system compared with laboratory reference methods traceable to a definitive standard. Based on POCT12-A3, the Clinical Laboratory Standards Institute standard for hospitals about hospital glucose meter procedures and performance, and Parkes error grid clinical accuracy performance criteria, no clinically significant differences were observed due to patient condition or therapy, with 96.1% and 99.3% glucose results meeting the respective criteria. Stratified sensitivity and specificity analysis (10 mg/dL glucose intervals, 50-150 mg/dL) demonstrated high sensitivity (mean = 95.2%, SD = ± 0.02) and specificity (mean = 95. 8%, SD = ± 0.03). Monte Carlo simulation modeling of the study blood glucose monitoring system showed low probability of category 2 and category 3 insulin dosing error, category 2 = 2.3% (41/1,815) and category 3 = 1.8% (32/1,815), respectively. Patient trend analysis demonstrated 99.1% (223/225) concordance in characterizing hypoglycemic patients.

CONCLUSIONS
The multicomponent, clinical accuracy assessment algorithm demonstrated that the blood glucose monitoring system was acceptable for use in critically ill patient settings when compared to the central laboratory reference method. This clinical accuracy assessment algorithm is an effective tool for comprehensively assessing the validity of whole blood glucose measurement in critically ill patient care settings.This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal.