Percutaneous coronary intervention is routinely performed to treat occlusive coronary artery disease. Coronary perforation is a potential complication and can be treated with a stent graft. Current stent grafts are associated with high restenosis rates. We tested the safety and feasibility of biodegradable stent grafts in pig and rabbit models.
Stent grafts were examined in pig coronaries with repeated OCT imaging for 42 days. Novel biodegradable coatings were applied on a bare metal stent by either an electrospinning (ES) or dip coating (DC) method. A completely biodegradable system was made by ES coating a magnesium-based stent. A commercially available stent graft served as a control. ES devices showed less restenosis (44.3 ± 8.8 vs. 59.1 ± 11.1% in controls, p < 0.05) and smaller reduction in minimum lumen area (44.3 ± 13.4 vs. 64.4 ± 13.6% in controls, p < 0.05) at day 42. DC devices occluded during follow-up. ES devices showed recanalization through the graft wall at day 42. Feasibility of the ES and DC devices was evaluated in pig coronary aneurysms and rabbit aortic perforation models and sealed aneurysms and perforations without complications.
Recanalization of the graft wall improves biocompatibility. Biodegradable stent grafts may present an alternative to permanent implants by showing reduced restenosis at day 42.
© 2020 S. Karger AG, Basel.
Related Posts
Advertisement
Meeting Coverage
- ACC 2020The American College of Cardiology decided to cancel ACC.20/WCC due to COVID-19, which was scheduled to take place March 28-30 in Chicago. However, ACC.20/WCC Virtual Meeting continues to release cutting edge science and practice changing updates for cardiovascular professionals on demand and free through June 2020.
- ENDO: 2020ENDO 2020 Annual Conference has been canceled due to COVID-19. Here are highlights of emerging data that has still been released. Keep an eye out for ENDO Online 2020, which will take place from June 8 to 22.