Antiβ2glycoprotein I (aβ2GPI) and anticardiolipin (aCL) IgG/IgM show differences in positive/negative agreement and titers between solid phase platforms. Method specific semiquantitative categorization of titers could improve and harmonize the interpretation across platforms.
To evaluate the traditionally 40/80 units thresholds used for aCL and aβ2GPI for categorization into moderate/high positivity with different analytical systems, and to compare with alternative thresholds.
aCL and aβ2GPI thresholds were calculated for two automated systems (chemiluminescent immunoassay (CLIA) and multiplex flow immunoassay (MFI)) by ROC-curve analysis on 1108 patient samples, including patients with and without APS, and confirmed on a second population (n=279). Alternatively, regression analysis on diluted standard material was applied to identify thresholds. Thresholds were compared to 40/80 threshold measured by an enzyme linked immunosorbent assay (ELISA). Additionally, likelihood ratios (LR) were calculated.
Threshold levels of 40/80 units show poor agreement between ELISA and automated platforms for classification into low/moderate/high positivity, especially for aCL/aβ2GPI IgG. Agreement for semiquantitative interpretation of aPL IgG between ELISA and CLIA/MFI improves with alternative thresholds. LR for aPL IgG increase for thrombotic and obstetric APS based on 40/80 thresholds for ELISA and adapted thresholds for the other systems, but not for IgM.
Use of 40/80 units as medium/high thresholds is acceptable for aCL/aβ2GPI IgG ELISA, but not for CLIA and MFI. Alternative semiquantitative thresholds for non-ELISA platforms can be determined by a clinical approach or by using monoclonal antibodies. Semiquantitative reporting of aPL IgM has less impact on increasing probability for APS.

This article is protected by copyright. All rights reserved.

Author